REXEED Dialyzers by Asahi
REXEED Dialyzers
Asahi has achieved a new level of overall performance in dialyzer design by combining the new polysulfone membrane REXBRANE with a new jacket design allowing for optimum flow dynamics. Incorporated on the inner surface is Asahi's unique hydrophilic gel layer technology. This minimizes blood membrane interaction and plays a crucial role in anticoagulation requirements. All has come together to create the new REXEED S (single use) and REXEED R (multiple-use) dialyzer.
A New Level of Overall Performance
  • Highest performance dialyzer in the US
  • Improved anticoagulation / less heparin
  • Available in four sizes:
    2.5m2 / 2.1m2 / 1.8m2 / 1.5m2
  • 2.5 m2 - highest performance dialyzer offered today
  • 1.5 m2 is ideal for Acute Patients
    Single and Multiple use
  • Gamma sterilized (No ETO residuals )
  • 2.5m2 - Alternative to “Tandem Dialysis”
  • High middle molecule removal
  • Next generation polysulfone membrane REXBRANE
  • Exceptional urea, creatinine and phosphate clearances

Improved Anticoagulation
Asahi’s unique membrane technology contributes to improved anticoagulation. Asahi’s hydrophilic gel layer helps to reduce the interaction between blood components and the membrane surface as illustrated below.

REXEED Dialyzers Improved Anticoagulation Hydrophilic Gel Layer

This reduces the potential for platelet activation and protein adhesion to the membrane. REXEED dialyzers frequently require less heparin than other dialyzers (graph below).

REXEED and Heparin usage

Improved Flow Dynamics
The improved flow dynamics of REXEED are provided by Asahi’s new high performance jacket design that provides a uniform dialysate flow.
The waved fiber design in REXEED provides smooth blood flow and improved dailysate penetration into the middle of the bundle for ideal dialysate / fiber contact and ideal small molecule clearance.

REXEED Dialyzer Improved Flow Dynamics

The ideal flow pattern is shown as a symmetrical sharp peak (above) that suggests quick dialysate turnover leading to exceptional small molecule clearance.

Balanced and Comprehensive
Small Molecule Clearance

REXEED has been developed to provide well balanced and comprehensive small and middle molecule clearance while minimizing albumin loss. This is demonstrated by the triangular graph (below).

REXEED Dialyzers


REXEED Dialyzers
REXEED Dialyzer Specifications
Dialyzer
Size
M2
Flux
Sterilization
Use
Membrane
KOA
Package
REXEED 15
1.5
Hi-Flux
Gamma
Single / Re-Use
Polysulfone (Rexbrane)
1232
12 / Case
REXEED 18
1.8
Hi-Flux
Gamma
Single / Re-Use
Polysulfone (Rexbrane)
1483
12 / Case
REXEED 21
2.1
Hi-Flux
Gamma
Single / Re-Use
Polysulfone (Rexbrane)
1683
12 / Case
REXEED 25
2.5
Hi-Flux
Gamma
Single / Re-Use
Polysulfone (Rexbrane)
1881
12 / Case
 
REXEED Dialyzers

REXEED Dialyzer Performance (in vitro)

REXEED-15
REXEED-18
Qd=500ml/min
Qd=800
ml/min
Qd=500ml/min
Qd=800
ml/min
Clearance
mL/min
Qb=200
mL/min
Qb=300
mL/min
Qb=400
mL/min
Qb=400
mL/min
Qb=200
mL/min
Qb=300
mL/min
Qb=400
mL/min
Qb=400
mL/min
Urea
196
272
317
352
198
280
330
366
Creatinine
191
255
296
326
194
265
309
341
Phosphate
185
238
272
301
190
250
289
320
Vitamin B12
142
168
179
199
152
183
197
219
KOA (Urea)
1232
1483
UFR
(mL/hr/mmHG)
63
71
REXEED-21
REXEED-25
Qd=500ml/min
Qd=800
ml/min
Qd=500ml/min
Qd=800
ml/min
Clearance
mL/min
Qb=200
mL/min
Qb=300
mL/min
Qb=400
mL/min
Qb=400
mL/min
Qb=200
mL/min
Qb=300
mL/min
Qb=400
mL/min
Qb=400
mL/min
Urea
199
284
340
374
199
287
350
380
Creatinine
195
272
317
352
196
280
330
364
Phosphate
192
257
299
334
194
260
303
346
Vitamin B12
160
194
212
236
163
199
225
251
KOA (Urea)
1683
1881
UFR
(mL/hr/mmHG)
74
80